Recall Z-1123-2012 ยท DXT Injector And Syringe, Angiographic

Mallinckrodt Inc

Coiled 60" Low Pressure Extension Tube 400 PSI Max; packaged in Plastic and Tyvec pouch, with 100 pouches per case; Liebel-Flarsheim Company, Cincinnati, OH 45237. Made in Mexico. Sterile, The purpose of this product is the injection of contrast media by means of power injector system during a diagnostic procedure.

DateFeb 29, 2012 (initiated Feb 10, 2012)
StatusTerminated
ReasonThe machine used to package the product was missing a tool, which could prevent the packaging from sealing properly, causing the sterility of the product to be compromised.
Root causePackaging process control
ActionCovidien sent an Urgent Device Recall letter dated February 10, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product immediately. Customers were to use the attached business response form to report the quantity of each lot that is currently in their stock. If customers did not have any of these lots in stock, they should indicate by checking the appropriate box on the business report form. When completed, customers should email the completed form to
DistributionWorldwide Distribution - USA including Distributed worldwide: USA including the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY, Puert
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K882329L-F COILED EXTENSION TUBING
Mallinckrodt Group, Inc., Sep 2, 1988