ClearanceBrief
Recall Z-1194-2012 ยท DHR System, Test, Rheumatoid Factor

Beckman Coulter Inc.

SYNCHRON Rheumatoid Factor (RF) Reagent, Part Number: 475136 A rheumatoid factor immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the rheumatoid factor (antibodies to immunoglobulins) in serum, other body fluids, and tissues.

DateMar 8, 2012 (initiated Jun 20, 2011)
StatusTerminated
ReasonThe recall was initiated because Beckman has confirmed that the all in date lots of Synchron Rheumatoid Factor (RF) Reagent might generate false positive results or calibration failures.
Root causeNonconforming Material/Component
ActionBeckman Coulter sent an Urgent Field Safety Notice/Product Correction letter dated June 20, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were notified that Beckman Coulter is developing a new RF assay. In the meantime due to current issues related to shelf life and low end sensitivity, Beckman has decided to take the product off the market effective immediately. Beckman will keep the customers updated on their progress and advise when the new assay becomes available. Customers were instructed to compl
Quantity22,081 units total (12,462 units in US)
DistributionWorldwide Distribution - USA (nationwide) and worldwide to foreign countries: Australia, Belgium, China, Costa Rica, Czech Republic, Finland, France, French Polynesia, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Lithuania, Macao, Mexico, Netherlands, New Zealan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K971788SYNCHRON SYSTEMS RHEUMATOID FACTOR (RF) REAGENT, SYNCHRON CX SYSTEMS RF CALIBRATOR
Beckman Instruments, Inc., Jun 3, 1997