Recall Z-1099-2012 ยท HKI Camera, Ophthalmic, Ac-Powered

Clarity Medical Systems Inc

Retcam3 & Retcam Shuttle with version 6.0 software. Clarity Medical Systems, Inc. 5775 W. Las Positas Blvd, Suite 200 Pleasanton, CA 94588 General ophthalmic imaging including retinal, corneal and external imaging. Photodocumentation of pediatric ocular diseases including retinopathy or prematurity (ROP). Screening of Type-2 re-threshold retinopathy of prematurity or treatment requiring ROP.

DateFeb 24, 2012 (initiated Feb 2, 2012)
StatusTerminated
ReasonStill images extracted from video recordings may be mislabeled as to right and left eye.
Root causeSoftware design
ActionClarity Medical Systems, Inc. sent a Safety Alert letter dated February 7, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customer was instructed the following : 1. Do not use the video mode until the software has been updated. 2. Install the software update using the enclosed CD. The instructions for installation are also enclosed. This software update will correct the anomaly. 3. If you have extracted still images from video recordings, review all potentially affected patient files to determine if any
Quantity33 RetCam Shuttles and 20 RetCam 3 - total 53.
DistributionWorldwide Distribution -- USA (nationwide) including the states of TN, MO, TX, FL, NJ and Puerto Rico and countries of Italy, Portugal, Czech Republic, Thailand, Chile, Oman, Hong Kong, Russia, Philippines, Sweden, Finland, Ecuador, Australia and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081858RETCAM 3 OPHTHALMIC IMAGING SYSTEM
Clarity Medical Systems, Nov 5, 2008
K090326RET CAM II, RETCAM 3 AND RETCAM SHUTTLE AND PORTABLE OPHTHALMIC IMAGING SYSTEMS
Clarity Medical Systems, Nov 25, 2009
K102859RETCAM 3 OPTHALMIC IMAGING SYSTEM
Clarity Medical Systems, Apr 20, 2011