Recall Z-1202-2012 ยท NAY System, Surgical, Computer Controlled Instrument
Intuitive Surgical, Inc.
da Vinci S. Surgical System, da Vinci Si and da Vinci Si-e Surgical Systems. Intuitive Surgical, Inc. Sunnyvale, CA 94086. The Intuitive Surgical Endoscopic instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
| Date | Mar 13, 2012 (initiated Feb 6, 2012) |
| Status | Terminated |
| Reason | The holding brake, a component of the da Vinci S Surgical System, may allow passive uncontrolled motion due to gravity during specific power-off conditions. |
| Root cause | Device Design |
| Action | Intuitive Surgical sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated February 3, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter provided new instructions for use. Additionally, a Device Correction Response Form was enclosed for customers to complete and return via fax to 408-523-1390. Contact Intuitive Surgical Customer Service at 1-800-876-1310 for questions regarding this notice. |
| Quantity | 1966 systems are in commerce at hospital locations. |
| Distribution | Worldwide Distribution. |