Recall Z-1202-2012 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

da Vinci S. Surgical System, da Vinci Si and da Vinci Si-e Surgical Systems. Intuitive Surgical, Inc. Sunnyvale, CA 94086. The Intuitive Surgical Endoscopic instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.

DateMar 13, 2012 (initiated Feb 6, 2012)
StatusTerminated
ReasonThe holding brake, a component of the da Vinci S Surgical System, may allow passive uncontrolled motion due to gravity during specific power-off conditions.
Root causeDevice Design
ActionIntuitive Surgical sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated February 3, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter provided new instructions for use. Additionally, a Device Correction Response Form was enclosed for customers to complete and return via fax to 408-523-1390. Contact Intuitive Surgical Customer Service at 1-800-876-1310 for questions regarding this notice.
Quantity1966 systems are in commerce at hospital locations.
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063220DA VINCI S SURGICAL SYSTEM-V1.1, MODEL IS2000
Intuitive Surgical, Inc., Dec 1, 2006
K081137INTUITIVE SURGICAL DA VINCI SI SURGICAL SYSTEM: MODEL IS3000
Intuitive Surgical, Inc., Feb 18, 2009