Recall Z-1127-2012 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

Siemens syngo.plaza with software version VA20B_HF04 Product Usage: Radiological image processing system

DateFeb 29, 2012 (initiated Jan 20, 2012)
StatusTerminated
ReasonThere is a potential malfunction when using syngo.plaza VA20B_HF04 with USB dongle and archive license. In syngo.plaza VA20B_HF04, there is a scenario where data in Short-Term Storage (STS) could reaming unprotected and not archived. this can happen when an archive server with a USB dongle-based license is enabled with auto archiving. If the disk fill level is reached which triggers auto deleti
Root causeSoftware design
ActionSiemens sent a Customer Safety Advisory Notice letter dated January 20, 2012 to all affected consignees. The letter identified the affected product, problem, steps to take to avoid this issue, and actions to be taken. Customers were instructed to contact their local Siemens Service to plan for their software update and ensure that this safety advisory is placed in the system's instruction for use.
Quantity7
DistributionNationwide Distribution (USA) - CA, FL, GA, MO, NY and WA.