Recall Z-1134-2012 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Philips Healthcare Inc.

Philips IntelliVue Patient Monitors: The monitor is intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates in health care facilities. The monitor is to be used by trained health care professionals. Physiological Monitoring, Patient Monitor , Arrhythmia detector and alarm (including ST -segment measurem

DateMar 1, 2012 (initiated Feb 9, 2012)
StatusTerminated
ReasonCentral station monitors may not alarm (either visual or audible) when an alarm is announced at the affected patient monitor
Root causeSoftware change control
ActionPhilips Healthcare sent a Urgent Medical Device Correction letter dated February 7, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Philips has recently received reports that, under certain circumstances, alarms announced at the patient monitor are not announced (either visual or audible) at the central station. This issue affects the following IntelliVue patient monitor models with SW rev. H.xx.xx MP20 (M8001A), MP30 (M8002A), MP40 (M8003A), MP50 (M8004A), MP60 (M8005A), MP70 (M8007A), MP80 (M8008A), MP90 (M8
Quantity83,534 units
DistributionWorldwide Distribution -- USA (nationwide) and the countries of ALBANIA, ANGOLA, ANTILLES, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BOLIVIA, BOTSWANA, BRAZIL, BYELORUSSIA, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, DENMARK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110622INTELLIVUE PATIENT MONITOR
Philips Medizinsysteme Boeblingen GmbH, Cardiac AN, Mar 31, 2011