ClearanceBrief
Recall Z-1372-2012 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Inc.

Access Immunoassay Systems Assay Protocol File (APF) Software versions for Access. Part Number: A19889, Access APF Wash Buffer REF 81907, Part Number: A45144, Access APF Wash Buffer II REF A16792. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentration found in human body fluids.

DateApr 6, 2012 (initiated Sep 11, 2008)
StatusTerminated
ReasonThe recall was initiated because Beckman confirmed a customer report that a timing conflict can occur while aspirating samples for the Access Total BhCG assay.
Root causeSoftware design
ActionBeckman Coulter sent a "PRODUCT CORRECTIVE ACTION" letter dated September 24, 2008 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers were instructed to share this information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. Customers were instructed to complete and return the enclosed response form. Customers with questions regarding this notification were instructed to contact Technical Support at (800) 854-3633 or their local Beckman Coulter representa
Quantity2948 units total (2762 units in the US)
DistributionWorldwide Distribution-USA (nationwide) and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K922823ACCESS THEOPHYLLINE IMMUNOASSAY REAGENTS
Bio-Rad Laboratories, Inc., Aug 5, 1992