Recall Z-1215-2012 ยท LFL Instrument, Ultrasonic Surgical

Integra LifeSciences Corp

Integra CUSA EXcel Sterile Tips Rx Only Single Use Only Product Usage: For usage in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable.

DateMar 15, 2012 (initiated Oct 14, 2011)
StatusTerminated
ReasonA small number of non-conforming CUSA Excel 23 KHz Standard Tip and flue (item Number C4601S) were released for distribution that did not meet Integra's in-process requirements for release. It was discovered during in-process inspection that the max power reading exceeded the specification of 80 watts, causing the stroke range of the tip to exceed it required range by up a value up to 0.002 inches
Root causeRelease of Material/Component prior to receiving test results
ActionIntegra sent an Urgent Product Recall Notification letter dated October 14, 2011 to all affected customers via US Mail. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine their inventory, discontinue use, quarantine and return any available affected lot numbers. Customers were instructed to distribute this communication to any customers to whom they have shipped the affected product. Customers were advise to complete the enclosed Recall Acknowledgement and Return Form and return it as soon as possible to Integra Customer Servic
Quantity24
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K981262CUSA EXCEL ULTRASONIC SURGICAL ASPIRATOR SYSTEM
Valleylab, Inc., Jul 6, 1998