Recall Z-1066-2012 ยท IYE Accelerator, Linear, Medical

Elekta, Inc.

Elekta Synergy XVI 4.5 and R4.6 To be used as part of radiation therapy treatment process.

DateFeb 21, 2012 (initiated Jan 26, 2012)
StatusTerminated
ReasonIf a user, using an on-line correction protocol, clicks the accept correction button, an approval dialog box appears. If cancel is selected, the table correction values return to their initial values, but the image panes continue to display the changed position.
Root causeSoftware design
ActionELEKTA sent an Important Field Safety Notice (20001507042) dated January 2012 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. For questions regarding this recall call 770-670-2548
Quantity96 units
DistributionNationwide Distribution including AR, AZ, CA, CT, FL, GA, ID, IL, LA MD, MA, MI, MN, MO, NJ, NY, NC, ND, OH, OK, OR, PA, RI, TN, VA, WA, WI, and Puerto Rico
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100115XVI R4.5
Elekta , Ltd., Mar 10, 2010