Recall Z-1273-2012 ยท JHX Fluorometric Method, Cpk Or Isoenzymes
Siemens Healthcare Diagnostics
IMMULITE 2000/IMMULITE 2000 XPi CK-MB L2KMB --- Catalog number: L2KMB2 (200 tests), L2KMB6 (600 tests) --- Test code: CMB Color: Dark Blue. For in vitro diagnostic use with the IMMULITE 2000 Analyzers --- for the quantitative measurement of creatine kinase isoenzyme MB (CK-MB) in heparinized plasma or serum, as an aid in patient management and the assessment of prognosis of myocardial infarction.
| Date | Mar 21, 2012 (initiated Jan 25, 2012) |
| Status | Terminated |
| Reason | Siemens Healthcare Diagnostics has observed that the median for the reference range with current kit lots of the CK-MB assay differs from the guidelines published in the Instructions for Use (IFU). This difference applies to all IMMULITE platforms. |
| Root cause | Nonconforming Material/Component |
| Action | Siemens sent an "URGENT DEVICE RECALL" letter dated January 2012 to all affected customers. The letter identifies the product , problem, and actions to be taken by the customers. Siemens recommends that customers reestablish their own medians and reference ranges, in line with the recommendations made in the IFU for these products. A Field Correction Effectiveness Check form was included for customers to complete and return to the firm. Customers with questions or who need additional information should contact their Siemens Technical Solutions Center. |
| Distribution | Worldwide Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K004002 | IMMULITE CK-MB, MODEL LKMB1 (100 TESTS), LKMB5 (500 TESTS); IMMULITE 2000 CK-MB, MODEL L2KMB2 (200 TESTS), L2KMB6 (600) Diagnostic Products Corp., Feb 14, 2001 |