Recall Z-1116-2012 ยท KHO Fluorometer, For Clinical Use

PerkinElmer Life and Analytical Sciences, Wallac, OY

Perkin Elmer Genetic Screening Processor (GSP) Product code: 2021-0010. It is intended for in vitro quantitative / qualitative determination of analytes in body fluids.

DateFeb 28, 2012 (initiated Jan 9, 2012)
StatusTerminated
ReasonGSP instruments with defective washer modules may cause a false negative screening result being reported for T4, TSH and IRT. For 17-OHP, a false positive result may be generated if the failure occurs for a normal sample.
Root causeNonconforming Material/Component
ActionPerkin Elmer issued an "Urgent Medical Device Recall" letter dated 1/25/12 to all customers and informed them about the possible issue and requested permission for service personnel to check their instrument. The customers were instructed that the instrument could not be used and the washer module required replacement. Customers were told to expect to be contacted by the firm's local representative for further details.
Quantity10 units
DistributionWorldwide Distribution - USA, including the states of Florida, Ohio, Oregon, Texas and the countries of Brazil, Belgium, France, Germany, and the UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090846GSP INSTRUMENT AND GSP NEONATAL HTSH KIT
Perkinelmer, Inc., Sep 3, 2009