Recall Z-1212-2012 ยท IZI System, X-Ray, Angiographic

Philips Healthcare Inc.

Philips Healthcare MD Eleva system. Multi-functional fluoroscopy, radiography and angiography systems that can perform more specialized interventional applications

DateMar 29, 2012 (initiated Nov 19, 2011)
StatusTerminated
ReasonIt was discovered there was a software error where the air karma rate is not displayed correctly on the user interface.
Root causeSoftware design
ActionA field correction action will be undertaken to correct the software error in affected systems worldwide. Your proposed action to bring the product into compliance with regulatory requirements is to institute a worldwide field correction to replace the software error in affected systems by updating the current software to version R6.1.2. The timeframe for completion of this field action will be 6 months post field action publication. Preventive actions: The software has been corrected moving forward in subsequent releases for these products. For further questions please call (978) 787-150
Quantity92 ( 24 Eleva 11s, and 68 Eleva FDs )
DistributionWorldwide Distribution - USA (nationwide)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.