Recall Z-1015-2012 ยท FRN Pump, Infusion

B. Braun Medical, Inc.

B. Braun Infusomat Space Pump - US Version Infusion pump. Model number 8713050U.

DateFeb 29, 2012 (initiated Jul 12, 2011)
StatusTerminated
ReasonIf the IV set anti free flow clip is incorrectly inserted into the pump and the pump door forced closed, the clip catch on the inside of the pump door may break. If the clip catch is broken and the door opened, free flow protection is ensured. If IV set is then removed from pump without closing the roller clamp, the clip may not re-engage the tubing, leaving an open fluid path with free flow pote
Root causeDevice Design
ActionB Braun sent a letter on August 16, 2011, to their customers, that detailed the proposed G03 implementation in response to reported customer complaints. The software upgrade with the door jam was agreed upon and the eight hospitals were upgraded in November 2011.
Quantity5769
DistributionProduct was distributed to hospitals in CA and MO.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062700INFUSOMAT SPACE VOLUMETRIC INFUSION PUMP SYSTEM
B.Braun Medical, Inc., Oct 26, 2006
K083689INFUSOMAT SPACE VOLUMETRIC INFUSION PUMP SYSTEM
B.Braun Medical, Inc., Jul 1, 2009