Recall Z-1075-2012 ยท GTY Antigens, All Groups, Streptococcus Spp.

Alere San Diego

Strep A Twist Rapid Test Cassette (K023766), IST-502T Twist Cassettes. Brand Name (Catalog Number): ACON Rapid Device Strep A Twist (IST-502T); CONSULT Diagnostics Strep A Twist, PSS (5005); Medi-Lab Performance Strep A Test-Twist, McKesson (32-502T); MooreBrand Strep A Twist Cassette, Moore (82793); and SP Strep A Rapid Twist Cassette Test, Cardinal Health (B1077-29). Type of Packaging (ie. box

DateFeb 22, 2012 (initiated Sep 1, 2011)
StatusTerminated
ReasonAlere San Diego is expanding the scope of the voluntary recall of the Strep A Rapid Test Cassette and Twist devices from the single lot that was recalled in January 12, 2011 to all lots of product that were on the market within the expiry period.
Root causeProcess change control
ActionThe firm, Alere, sent an "URGENT MEDICAL DEVICE RECALL" letter dated September 1, 2011 to its customers. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to retain the recall letter for their records as documentation that they have received the recall information. Customers were instructed to follow the directions and return the attached Verification Form. Customers were instructed to use the table in the Verification Form to document all inventory at any of their inventory locations. Customers may use mult
Quantity2,797,600 devices (228 lots)
DistributionWorldwide distribution: USA (nationwide) and countries of : Austria, Canada, Cyprus, Denmark, Finland, Germany, Greece, Hungary, Iceland, Italy, Poland, Portugal, Romania, Slovenia, Spain, Switzerland, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023766ACON STREP A TWIST TEST DEVICE
Acon Laboratories Co., Feb 10, 2003