ClearanceBrief
Recall Z-1028-2012 ยท KWA Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)

Wright Medical Technology Inc

DYNASTY(R) BIOFOAM(R) SHELL, SIZE: 70mm GROUP J, REF DSBFGJ70, Rx ONLY, STERILE R, Wright Medical Technology, Inc., hip prosthesis acetabular component. The instrument is used as a hip acetabular shell.

DateFeb 14, 2012 (initiated Jan 3, 2012)
StatusTerminated
ReasonThe possibility exists that particulate debris from the packaging spacer could be unintentionally implanted into a patient and could potentially lead to complications that might necessitate a revision surgery.
Root causePackaging
ActionWright Medical sent an "URGENT PRODUCT RECALL NOTIFICATION" letter dated December 27, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to check their inventory for the affected product and to return any affected product to the firm. A Product Recall Verification Form was enclosed for customers to complete and return via fax to 901-867-7401. Contact Customer Service at 1-800-238-7117 for questions regarding this recall.
Quantity23 units
DistributionWorldwide Distribution-USA (nationwide) and the countries of Australia, Brazil, Canada, Denmark, and Italy.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082924DYNASTY POROUS ACETABULAR SHELL, DYNASTY POLYETHYLENE ACETABULAR LINER, DYNASTY METAL ACETABULAR LINER
Wrightmedicaltechnologyinc, Mar 6, 2009