Recall Z-1038-2012 ยท LDF Electrode, Pacemaker, Temporary

European Custom Manufacturing B.V.

Temporary Pacing Wire, BIPOLAR, Models BM, V, VE and VF, Rx Only, STERILE/EO, Single use only -- Mce (MEDICAL CONCEPTS EUROPE) --- Manufacturer ECM B.V. Oost-Om 54, P.O. Box 53, NL 5420 AB GEMERT --- Shipping or Unit Package: 20 individual units packaged per box with 6 boxes per case. --- Device Classification Name: Electrode, pacemaker, temporary. ---The device is supplied sterile and intended f

DateFeb 15, 2012 (initiated Apr 18, 2011)
StatusTerminated
ReasonThe crimping operation during product production was not performed properly. As such, the pacing wire may separate from the connection pin, which causes non-functionality.
Root causeProcess design
ActionECM sent an "ECMbv RECALL OF MCE TEMPORARY PACING WIRES" letter dated May 2, 2011 to all affected customers. Customers were also informed of a potential problem via e-mail and/or phone between April 18 and 21, 2011. The letter describes the product, problem, and actions to be taken by the customers. Customers were instructed to place the affected product on hold and to return the product to the firm. For questions on this matter, contact ECM or MCE Customer Service at 1-877-886-1985.
Quantity2,120 individual units (106 boxes)
DistributionNationwide Distribution-including the states of AL, AR, CA, FL, GA, KY, NC, NV, NY, TX, VA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033858TEMPORARY BIPOLAR MYOCARDIAL PACING WIRE
European Custom Manufacturing BV, Jul 12, 2004