Recall Z-0876-2012 ยท FRN Pump, Infusion

CareFusion 303, Inc.

Alaris PC unit model 8015. The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System.

DateJan 26, 2012 (initiated Sep 13, 2011)
StatusTerminated
ReasonThe recall was initiated because Carefusion has identified a potential risk with a PLD component on the Alaris PC unit model 8015 power supply board. The PLD component used in production between June 21, 2011 and August 15, 2011 has a higher standby current resulting in the battery depleting prematurely.
Root causeNonconforming Material/Component
ActionCarefusion sent an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" letter dated September 13, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. A Customer Reply Card was enclosed for customers to complete and return to the firm. Customers were instructed to call Carefusion Support Center at (888) 562-6018, 7 am-5 pm (Pacific) for any recall related questions.
Quantity1039 units
DistributionWorldwide Distribution-USA (nationwide) including the states of AZ, CA, CT, FL, IA, IL, MI, MO, NJ, NY, OH, SC, SD, and TX. and the country of Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051641ALARIS SYSTEM PC UNIT, MODEL 8001
Cardinal Health, Alaris Products, Oct 20, 2005