Recall Z-0919-2012 ยท FTL Mesh, Surgical, Polymeric

Novomedicus Llc

The Artificial Bur Patch is a Temporary Fascial Prothesis. For Single Patient Use only. Use with Negative Pressure Wound Dressing. Contents: Large Bur Patch (Sterile: EO), 20cm x 40cm Loop Sheet, 20cm x 40cm Hook Sheet. Sterile: Contents are sterile for maximum two years unless enclosed package has been opened or damaged. Manufactured For: NOVOMEDICUS, LLC P.O. Box 2105 Nokomis, FL 34274. Used f

DateJan 27, 2012 (initiated Jun 22, 2011)
StatusTerminated
ReasonOn 06/22/2011 Novomedicus, FL initiated a recall of the Artificial Bur Hypopack Lot 1010.1, 1102.1, 1103.1 and 5643.2. Cytotoxicity testing revealed a biocompatibility issue. The test results did not represent a health risk.
Root causeProcess change control
ActionNovomedicus sent a Medical Device Correction letter dated June 22, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to destroy the inventory, or return the inventory to Novomedicus. If customers decided to return the remaining inventory, Novomedicus would provide TWO replacement kits for each one returned at no additional cost. Customers were asked to complete the attached Return Response form and return to NovoMedicus using the self-addressed and stamped envelope included. For quest
Quantity10 patches
DistributionWordwide Distribution - USA including IL, AZ, PA, CT, MO, WI, NY, TX, OR, ME, FL, VA, and MS and the countries of Mexico and Qatar.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093166WITTMANN PATCH
Aperta, LLC, Dec 7, 2009