Recall Z-1132-2012 ยท INX Bed, Air Fluidized
Hill-Rom, Inc.
Clinitron Rite Hite Bed headboard and foot side rails. The Clinitron Rite Hite Air Fluidized Therapy Unit is a patient management system that combines air fluidized therapy and low airloss therapy technologies together on a low-height frame. The lower body section provides air fluidized therapy, and the upper body section provides low airloss therapy.
| Date | Mar 1, 2012 (initiated Feb 3, 2012) |
| Status | Terminated |
| Reason | The firm identified in October 2011 that the Clinitron Rite Hite Bed and Clinitron At Home Bed headboards failed to meet the safety standard force requirements. The firm identified in October 2011 that the Clinitron Rite Hite Bed and Clinitron At Home Bed outer rail, which is extruded Aluminum tubing, is intended to be one piece joined together by a weld at the lower center. The parts receiv |
| Root cause | Nonconforming Material/Component |
| Action | Hill-Rom, Inc. sent a Modification Notice Urgent Medical Device Correction Notification dated February 2, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The modification notice instructed all customers to find and replace all headboards and food side rails that do not meet design intent. All technicians will inspect all affected product in the service centers and all at home and rite hite beds for an "x" or "L" stamp. The technicians will inspect the beds that are still in use during scheduled maintenance o |
| Quantity | Foot Siderail = 70; Headboards = 103 |
| Distribution | Worldwide Distribution -- USA (nationwide) including the states of AL, AK, AZ, AR, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, WA, WI, WV, and WY. and the country of Canada. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K964223 | CLINITRON RITE-HITE Hill-Rom, Inc., May 28, 1997 |