Recall Z-1448-2012 ยท FKX System, Peritoneal, Automatic Delivery

Fresenius Medical Care Holdings, Inc.

Fresenius Liberty Cycler Software, User Manuals, and Handi-Guides for acute and chronic peritoneal dialysis. Software versions earlier than 2.8. The Fresenius Liberty Cycler is intended for acute or chronic peritoneal dialysis.

DateApr 24, 2012 (initiated Mar 30, 2012)
StatusTerminated
ReasonIncreased risk of intraperitoneal volume (IIPV), referred to as Overfill
Root causeSoftware design
ActionFresenius Medical Care sent a Urgent Product Information letter dated March 30, 2012, and April 3, 2012, to all affected customers. The letter was sent via Certified Mail, Return Receipt Requested. The letter identified the product the problem and the action needed to be taken by the customer. Recall to occur in 2 phases: Phase I: Fresenius Medical Care notified all clinics with patients using Liberty Cyclers with software versions earlier than 2.8 that these patients are receiving updated User Manuals and Handi-Guides. These new materials will include updated descriptions of the safety w
Quantity10,146 units
DistributionWorldwide Distribution -- USA (nationwide)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043363FRESENIUS LIBERTY CYCLER
Fresenius Medical Care North America, Mar 31, 2005