ClearanceBrief
Recall Z-0855-2012 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Zimmer Inc.

Natural-Knee II Durasul All-Poly Patella, Size 3, 10mm The Zimmer Natural-Knee II Durasul Patella product is a semi-constrained total knee prosthesis consisting of four anatomically designed components: the femoral, tibial, articular surface, and patellar prosthesis. The all-polyethylene patella is made from Durasul Highly Crosslinked Ultra High Molecular Weight Polyethylene. The N-K II Durasul p

DateJan 25, 2012 (initiated Dec 21, 2011)
StatusTerminated
ReasonThe firm has 16 reports of patella peg shear out of 109,386 distributed units (0.0137%). The failures have been reported where the pegs on the back side of the patella have sheared at the peg/patella interface. Analysis shows that the highly cross-linked polyethylene, Durasul, patella was shown to have a lower fracture fatigue strength than the Sulene counterpart and that the cement technique used
Root causeComponent design/selection
ActionZimmer sent an Urgent Device Recall letter dated December 20, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to stop distribution of the affected product and return all lots in inventory. Customers should provide a list of all hospitals and surgeons that have used the affected product by completing the User Facility and Health Care Professional Form and fax a copy to (574) 372-4265. If customers distributed the product further they should provide a copy of the recall letter to those cust
DistributionWorldwide Distribution - USA (nationwide) and the countries of Belgium, Germany, France, Austria, Sweden, Switzerland, South Africa, and the UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002335NATURAL-KNEE II SYSTEM-DURASUL TIBIAL INSERT AND DURASUL ALL-POLY PATELLA
Sulzer Orthopedics, Inc., Oct 25, 2000