Recall Z-0691-2012 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems, Inc. Oncology Systems

4D Integrated Treatment Console (4DITC) TrueBeam Linear Accelerator Varian Medical Systems, Palo Alto, CA. The 4DITC is designed to interface to the Varian Clinac Treatment Control System for the purpose of providing verification of treatment parameters, imaging and providing a record of treatment.. The True Beam is a medical linear accelerator which delivers X-ray and electron radiation for ster

DateJan 11, 2012 (initiated Dec 12, 2011)
StatusTerminated
ReasonVarian has discovered that on 4D Integrated Treatment Console and TrueBeam, the ad-hoc CIAO(Complete Irradiated Area Outline) calculation results in the CIAO generation that is different from the CIAO in RT Chart. The 4DITC algorithm adds dose for the control point at field edge, which should not be added according to the system specifications.
Root causeDevice Design
ActionVarian Medical Systems sent a Urgent Medical Device Notice dated December 22, 2011, to all affected customers via certified mail. The letter identified the product the problem and the action needed to be taken by the customer. The Letter will also be distributed to the Varian Sales, Marketing, and Service organizations, informing them of the recall. ***12-20-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.*** Canadian account letters were sent by Fed Ex. International Customer Support Managers were sent
Quantity4875 units
DistributionWorldwide Distribution -- USA (nationwide)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K0911324D INTEGRATED TREATMENT CONSOLE
Varian Medical Systems, Jun 17, 2009
K111106TRUEBEAM
Varian Medical Systems, Inc., Aug 18, 2011