Recall Z-0716-2012 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems, Inc. Oncology Systems

Varian brand Barcode Conical Collimator Verification (BCCV) v1.0, Reference/FSCA Identifier: CP-06873, Model Number: HB, distributed by and/or manufactured by: Varian Medical Systems Inc., Palo Alto, CA The 4D Integrated Treatment Console [4DITC] is designed to interface to the Varian Clinac Treatment Control System [TCS] for the purpose of providing verification of treatment parameters, imaging

DateJan 11, 2012 (initiated Dec 12, 2011)
StatusTerminated
ReasonWhen multiple conical collimator treatment plans are scheduled to be treated in a session, the Plan Label should be displayed on the BCCV dialog box should match the Plan Label on the 4DITC. However, the name of the plan is not being updated correctly.
Root causeDevice Design
ActionThe firm, VARIAN Medical Systems, sent an "URGENT MEDICAL DEVICE CORRECTION-URGENT FIELD SAFETY NOTICE" dated December 12, 2011, to all affected customers. The notice described the product, problem and user corrective action steps. The customers were provided to do the following: Do not use BCCV as a primary reference for identifying the correct plan to be treated; use caution when selecting an isocenter in a positioning system that is not interfaced to the 4DITC; any discrepancies in planned versus actual parameters should be brought to the attention of the medical physicist for investigation
Quantity9
DistributionWorldwide distribution:USA (nationwide) and countries of: Canada, Australia, Austria, Brazil, China, France, Georgia, Germany, Hong Kong, Hungary, India, Ireland, Italy, Japan, Latvia, Lithuania, Macau. Malaysia, Mexico, Netherlands, Singapore, Slovenia, South Korea, Spain, Sri Lanka, Switzerland,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K103394BARCODE CONICAL COLLIMATOR VERIFICATION (BCCV)
Varian Medical Systems, Oncology Systems, Jan 12, 2011