Recall Z-0984-2012 ยท LEG Enzyme Immunoassay, Valproic Acid

Ortho-Clinical Diagnostics

VITROS Valproic Acid (VALP) Reagent, REF/Catalog No. 6801710 --- Ortho- Clinical Diagnostics Rochester, NY USA -- Ortho- Clinical Diagnostics High Wycombe, UK --- An in vitro diagnostic. For in vitro diagnostic use only. Used to quantitatively measure valproic acid (VALP) concentration in human serum and plasma.

DateFeb 8, 2012 (initiated May 10, 2011)
StatusTerminated
ReasonResults using VITROS Chemistry Products VALP Reagent were negatively biased compared to other manufacturers. OCD confirmed that patient results were also affected by this issue.
Root causeManufacturing material removal
ActionAll domestic end-user consignees were initially sent "Important Product Correction Notification" recall communications (dated 5/10/11) by Federal Express overnight mail on 5/10/11. The letters informed users of the issue, and advised them to discontinue using their VITROS VALP reagent until adjusted calibrator values are provided for GEN 14 of VITROS VALP Reagent. The adjusted calibrator values for VITROS VALP Reagent were sent to customers via Assay Data Disk (ADD) Data Release Version (DRV) 5632. Customers were to refer to the Technical Support Communication for details. Users were to co
QuantityApproximately 5800 units worldwide (domestic - 4549, foreign - 1287)
DistributionWorldwide Distribution -- US, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042476VITROS CHEMISTRY PRODUCTS VALP REAGENT; VITRO CHEMISTRY PRODUCTS CALIBRATOR KIT 12; VITROS CHEMISTRY PRODUCTS TDM PERFOS
Ortho-Clinical Diagnostics, Inc., Dec 1, 2004