Recall Z-0913-2012 ยท FSY Light, Surgical, Ceiling Mounted
Stryker Communications Corp
Visum LED Surgical Lights, Ceiling Mounted Surgical Light. The Visum LED surgical light is intended to illuminate the operative site during surgical procedures with high intensity light.
| Date | Jan 26, 2012 (initiated Nov 7, 2011) |
| Status | Terminated |
| Reason | Stryker has identified a potential structural issue with the Visum LED surgical light(s). During a complaint investigation, Stryker became aware that there was a possibility that incorrect assembly could result in the light head beginning to separate from the suspension arm. In the event the light separates and falls from the arm, it could lead to serious injury. |
| Root cause | Nonconforming Material/Component |
| Action | The firm sent "Urgent: Medical Device Correction" letters dated 11/07/2011 to their domestic and international customers. The letter informed customers of the remote risk of the light detaching from the arm and recommended continued use of the product. A representative will contact them about scheduling an appointment to inspect the lights. Contact Technical Services at 1-866-841-5663 for questions. |
| Quantity | 1143 Devices |
| Distribution | Worldwide Distribution - USA, and the countries of Australia, Canada, Hong Kong, Brazil, Sweden, India, Japan, Korea, Ecuador, Venezuela, Mexico, Denmark, Great Britain, and New Zealand. |