Recall Z-0913-2012 ยท FSY Light, Surgical, Ceiling Mounted

Stryker Communications Corp

Visum LED Surgical Lights, Ceiling Mounted Surgical Light. The Visum LED surgical light is intended to illuminate the operative site during surgical procedures with high intensity light.

DateJan 26, 2012 (initiated Nov 7, 2011)
StatusTerminated
ReasonStryker has identified a potential structural issue with the Visum LED surgical light(s). During a complaint investigation, Stryker became aware that there was a possibility that incorrect assembly could result in the light head beginning to separate from the suspension arm. In the event the light separates and falls from the arm, it could lead to serious injury.
Root causeNonconforming Material/Component
ActionThe firm sent "Urgent: Medical Device Correction" letters dated 11/07/2011 to their domestic and international customers. The letter informed customers of the remote risk of the light detaching from the arm and recommended continued use of the product. A representative will contact them about scheduling an appointment to inspect the lights. Contact Technical Services at 1-866-841-5663 for questions.
Quantity1143 Devices
DistributionWorldwide Distribution - USA, and the countries of Australia, Canada, Hong Kong, Brazil, Sweden, India, Japan, Korea, Ecuador, Venezuela, Mexico, Denmark, Great Britain, and New Zealand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031068STRYKER VISUM SURGICAL LIGHTING SYSTEM
Stryker Endoscopy, Aug 12, 2003
K060802VISUM LED SURGICAL LIGHTING SYSTEM
Stryker Communications Corp., Apr 7, 2006