Recall Z-0826-2012 ยท FKR Subsystem, Proportioning

NxStage Medical, Inc.

NXStage PureFlow SL SAK-303 - 2 dialysate sacks (SAK) with concentrate for use with PureFlow SL, Rx The NxStage PureFLow SL is an optional accessory to the NxStage System One that prepares dialysate for use during hemodialysis, as prescribed by the physician.

DateJan 20, 2012 (initiated Dec 14, 2011)
StatusTerminated
ReasonReport of high conductivity failure alarms with the PureFlow SL due to excess concentrate by weight in the product (SAK) dialysate Preparation System
Root causeEmployee error
ActionOn December 13, 2011, NxStage sent replacement product for overnight delivery and initiated phone contact with all consignees of the product . In addition, NxStage sent an "IMPORTANT MEDICAL DEVICE RECALL" notice dated December 14, 2011 to its consignees/customers. The notice describes the product, problem and actions to be taken. The customers were instructed to use SAKs from a different lot or premixed dialysate bags (RFPs), until their replacement SAKs arrived; set aside all SAKs from listed lot number; record their inventory; dispose of the SAKs following their normal practices, and comple
Quantity1,192 devices
DistributionNationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080919NXSTAGE PUREFLOW SL
Nxstage Medical, Inc., Oct 15, 2008
K111174NXSTAGE PUREFLOW SL
Nxstage Medical, Inc., Sep 19, 2011