Recall Z-0860-2012 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Siemens Medical Solutions USA, Inc

Siemens AXIOM Aristos FX Plus solid state x-ray imager. Image-intensified fluoroscopic x-ray system.

DateJan 25, 2012 (initiated Dec 12, 2011)
StatusTerminated
ReasonFixation bolts for the mounting of the X-ray tube holder became mixed with an incorrect bolt type.
Root causeProcess control
ActionSiemens Healthcare sent a "SAFETY ADVISORY NOTICE" dated December 12, 2011 to the affected customer. The letter identifies the product, problem, and actions to be taken by the customer. The firm prepared a modification, which will be provided to the affected customer under Update Instruction XP055/11/S. The firm will resolve this potential issue by preventively exchanging all bolts for the fixation of the X-ray tube holder of the concerned AXIOM Aristos FX Plus unit.
Quantity1
DistributionProduct distributed in one hospital in California.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061054AXIOM ARISTOS FX PLUS
Siemens Medical Solutions USA, Inc., May 9, 2006