Recall Z-0870-2012 ยท BTL Ventilator, Emergency, Powered (Resuscitator)

Impact Instrumentation, Inc.

Model 73X Uni-Vent Product Usage: Intended for use at mass casualty incidents.

DateNov 4, 2011
StatusTerminated
ReasonA piece of internal tubing in the 73X portable ventilator is of incorrect specifications for the device.
Root causeComponent design/selection
ActionImpact Instrumentation Inc. sent an URGENT Device Product Recall letter dated November 3, 2011 to all affected consignees. The letter identified the product, problem, instructions for product returns, and actions to be taken. The letter states that Impact will arrange for expedited return, at no charge to the customers of the affected units. The notice instructs customers to complete and return the Recall Tracking Form. If you need additional information, please call Impact Customer Service at (800) 969-0750
Quantity57 units
DistributionWorldwide Distribution - USA (nationwide)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051476UNI-VENT MODEL 73X
Impact Instrumentation, Inc., Oct 28, 2005