Recall Z-0647-2012 ยท FGE Stents, Drains And Dilators For The Biliary Ducts

Navilyst Medical, Inc

Exodus(TM) 8F (2.7 mm)/ 35 cm Standard Loop Biliary Drainage Catheter with Hydrophilic Coating, REF/ Catalog No. 10005, UPN/Product No. H965100051, STERILE --- Package/Box Units: Box of 5 eaches --- Made in USA: 2301 Centennial Blvd. Jeffersonville, IN 47130 -- Manufactured for: Navilyst Medical, Inc. 26 Forest Street Marlborough, MA 01752 --- The Exodus Drainage Catheter consists of a radiopaque

DateJan 11, 2012 (initiated Feb 4, 2011)
StatusTerminated
ReasonNavilyst Medical, Inc. determined that the Exodus Standard Loop Biliary Drainage Catheters exhibited a rate of reported failures (catheter tip fragmentation) higher than anticipated.
Root causeNonconforming Material/Component
ActionThe firm, Navilyst Medical, sent an "Urgent Medical Device Recall -Immediate Action Required" letter dated February 4, 2011, via Federal Express to the domestic consignees/customers and one foreign consignee/customer. Consignees/customers were instructed to: (1.) Immediately discontinue use and remove the recalled product from inventory; (2.) Segregate this product in a secure location for return to Navilyst Medical, Inc.; (3.) Immediately forward a copy of the recall notification to ALL sites to which they have distributed affected device (4.) promptly complete and return the Reply Verifi
DistributionWorldwide distribution: USA (nationwide) including states of: AK, AZ, CA, CO, FL, IL, KS, MI, MO, NC, NJ, NY, OH, OK, PA, PR and country of: South Africa.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093392MULTIPURPOSE, NEPHROSTOMY AND BILIARY DRAINAGE CATHETERS
Navilyst Medical, Inc., Dec 31, 2009