Recall Z-0868-2012 ยท DKB Calibrators, Drug Mixture

Lin-Zhi International Inc

Multi-Analyte Urine DAU Calibrator (15 mL HDPE bottle inside a cardboard box) For in vitro diagnostic use only. Lin-Zhi International, Inc.. 670 Almanor Ave. Sunnyvale, CA 94085 intended use: Drug abuse detection.

DateJan 25, 2012 (initiated Aug 23, 2011)
StatusTerminated
ReasonMulti Analyte Urine DAU Cutoff Calibrator Gives a decreased absorbance reading due to a loss of the BZO (oxazepam) analyte from the solution and may cause false positive readings.
Root causeNonconforming Material/Component
ActionThe firm, Lin-Zhi International, Inc., sent a notice by email dated August 23, 2011 to all customers. The notice described the product, problem and actions to be taken. Lin-Zhi will contact the customers to determine whether they want replacement BZO calibrators and controls immediately or if they are willing to wait a minimum of four weeks for a new lot of the Multi-Analyte Urine Cutoff Calibrators. The customers were instructed to contact Lin-Zhi with their replacement request. The notice also stated that the Lin-Zhi has plans to create more stringent QC standards and further R&D regardin
Quantity45 items
DistributionWorldwide distribution: USA (nationwide) including states of: AL, FL, GA, IN, NC, NH, and TX; and country of: Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051088MULTIPLE ANALYTE URINE DRUGS OF ABUSE CALIBRATORS AND CONTROLS
Lin-Zhi International, Inc., Jun 24, 2005