Recall Z-0869-2012 ยท MED Sterilant, Medical Devices

Steris Corporation Hopkins Facility

STERIS SYSTEM 1E Liquid Chemical Sterilant Processing System. Model P6500, 510 (k) Numbers: K090036 Cleared 4/05/2010; K101409 cleared 8/2/2010; K102462, cleared 9/2/2010; add to file 2/24/2011. The SYSTEM 1E Liquid Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reuseable critical and semi-critical heat sensitive medical devices,

DateJan 25, 2012 (initiated Dec 7, 2011)
StatusTerminated
ReasonSoftware deficiency. STERIS identified that water specifications were not being met at some of their customer sites. Critical parameters for initiating cycles are: water quality, water pressure and water temperature. Failure to meet these criteriae may lead to cycle aborts by the SYSTEM 1E Liquid Chemical Sterilant Processing System. The firm intends to perform a software field correction which
Root causeSoftware design
ActionThe firm, STERIS Corp. prepared an "URGENT VOLUNTARY FIELD CORRECTION NOTICE" letter dated December 8, 2011. The firm intends to send the letter to its customers upon STERIS' receipt of 510(k) clearance from FDA. The letter includes a description of the Product-SYSTEM Liquid Chemical Sterilant Processing System, problem and actions taken. The customers were instructed to in case of an incomplete or cancelled cycle, reprocess devices in the cycle following the directions in the Operator manual. The letter also states that STERIS Corporation will install updated software on the affected sys
Quantity3,882 units were distributed
DistributionWorldwide distribution: USA (nationwide) and countries of: Bahrain, Hong Kong, Japan and South Korea.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090036SYSTEM 1.5 STERILE PROCESSING SYSTEM
STERIS Corporation, Apr 5, 2010
K102462SYSTEM 1E LIQUID CHEMICAL STERILANT PROCESSING SYSTEM
STERIS Corporation, Sep 21, 2010