ClearanceBrief
Recall Z-0692-2012 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems, Inc. Oncology Systems

Varian High Energy Linear Accelerator, C-Series Clinac, Reference/FSCA Identifier: CP-06611; Models Numbers: 600C, 600CD, 6EX, DBX, 2100C, 2100CD, 2300CD, 21EX, 23EX, DMX, DHX versions 2.x through 6.x; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA. The Varian High Energy Linear Accelerator is intended to provide stereotactic radiosurgery and precision rad

DateJan 11, 2012 (initiated Nov 28, 2011)
StatusTerminated
ReasonAn event has been reported to Varian which entails excessive connector resistance, which caused the actual jaw positions to differ from the intended jaw positions without warning the operator.
Root causeDevice Design
ActionVarian Medical Systems sent an Urgent Medical Device Correction - Urgent Field Safety Notice dated November 28, 2011 ***Expanded to include product #2 on December 16, 2011***, to all affected customers. ***12-12-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.*** The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to establish a comprehensive quality assurance program before beginning treatments for the day, quality assurance tests mus
Quantity6158 units total
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112839VARIAN HIGH ENERGY LINEAR ACCELERATOR
Varian Medical Systems, Nov 14, 2011
K926321VARIAN C/D-SERIES CLINACS (CLINAC 600 C/D)
Varian Assoc., Inc., Aug 9, 1994