ClearanceBrief
Recall Z-0630-2012 ยท FPA Set, Administration, Intravascular

ConMed Corporation

STAT 2 Extension Set Gravity Flow Controller, 12 in. length, Catalog/REF No. S2-12, STERILE, ASSEMBLED IN MEXICO, CONMED CORPORATION, 525 FRENCH ROAD, UTICA, NY 13502 --- This is a disposable device with a sterile, nonpyrogenic fluid path. The STAT 2® Extension sets are for use in conjunction with commonly available I.V. administration sets for controlling the infusion of I.V. fluids int

DateJan 11, 2012 (initiated Apr 15, 2011)
StatusTerminated
ReasonConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined that certain units were assembled with an incorrect component which could result in the devices exhibiting a different flow rate and might reduce the ability of the user to regulate the device at low flow rates.
Root causeMixed-up of materials/components
ActionConMed Corporation sent Urgent Medical Device Recall Letters and Response Forms dated April 14, 2011, via UPS Priority Mail pouches to all domestic consignees. The letter identified the product, the problem, and the action to be taken by the consignee. The international consignees were sent Urgent Medical Device Recall Letters and Response Forms dated April 22, 2011, via FedEx International Priority. Users were instructed to immediately stop use of the device and return it to ConMed Corporation. --- RECALL EXPANSION: Consignees were also instructed to contact all of those organization wi
DistributionWorldwide Distribution - USA (nationwide) and the countries of Canada, Panama and India
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K964844CONMED: STAT 2 AND STAT 2 PUMPETTE EXTENSION SETS AND MASTERFLOW AND MASTERFLOW PUMPETTE PRIMARY SETS
Conmedcorp, Mar 24, 1997