Recall Z-0811-2012 ยท DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass

Medtronic Inc. Cardiac Rhythm Disease Management

Medtronic, DLP, DLP Malleable Single Stage Venous CannulaeRx Only, Sterilized Using Ethylene Oxide, Nonpyrogenic. Model # 91037. These cannulae are intended for collection of venous blood from the right side of the heart via the superior and inferior vena cava during cardiopulmonary bypass surgery up to six hours or less.

DateJan 18, 2012 (initiated Dec 14, 2011)
StatusTerminated
ReasonInstructions for Use for the DLP Single Stage Venous Cannulae are in the process of being updated. The current precaution statement referencing ventricular assist procedures will be identified as a contraindication in the updated Instructions for Use. DO NOT USE THIS PRODUCT FOR EXTENDED TERMS SUCH AS VENTRICULAR ASSIST PROCEDURES. Medtronic has received two reports of patient deaths due to p
Root causeLabeling design
ActionMedtronic sent an "URGENT MEDICAL DEVICE INFORMATION" letter dated December 2011 to all affected customers. The letter identifies the product, problem, and the action to be taken by the customers. Contract Medtronic CardioVascular LifeLine Technical Support at 1-877-526-7890 for questions related to this issue.
Quantity883
DistributionWorldwide Distribution-USA (nationwide) including Puerto Rico and the Virgin Islands and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Barbados, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Chile
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K854487VENOUS RETURN CANNULA W/CUFF 91039
Dlp, Inc., Feb 10, 1986