ClearanceBrief
Recall Z-0798-2012 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

Mevatron KDS-2; Siemens Healthcare Concord, CA 94520 Product Usage: PRIMUS and MEVATRON Linear Accelerators are intended to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.

DateJan 17, 2012 (initiated Dec 1, 2011)
StatusTerminated
ReasonBolts holding the trim weight may break causing the plates of the counterweight to become loose and fall into the cover.
Root causeNonconforming Material/Component
ActionSiemens sent a Customer Information Letter on December 1, 2011 to all affected customers. The letter idenfied what is the issue, when does the issue occur and what Siemens will do to address the issue. The letter recommend that the Customer Information letter be included in the Linear Accelerator System Owner manual chapter Safety Advisory Letters.
Quantity327 Active Devices - all models
DistributionWorldwide Distribution - USA (nationwide)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K882729MEVATRON, M 2 CLASS
Siemens Medical Laboratories, Inc., Sep 19, 1988