Recall Z-0516-2012 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems, Inc. Oncology Systems

Portal Vision, R-Arm with 4D Integrated Treatment Console; Varian Medical Systems. Palo Alto, CA. The Portal Vision device is used for verification of the treatment field and shielding block in relation to anatomical landmarks in radiotherapy treatment. Portal Vision will also allow for verification of the exit dose in radiotherapy treatment.

DateJan 11, 2012 (initiated Nov 23, 2011)
StatusTerminated
ReasonAn anomaly was discovered when using the 4DITC in combination with an R-Arm to acquire double exposure MV images.
Root causeSoftware design
ActionThe firm, Varian Medical Systems, sent an "URGENT MEDICAL DEVICE CORRECTION/URGENT FIELD SAFETY NOTICE" dated December 1, 2011 to all affected sites via Certified mail, return receipt. Canadian accounts were sent by Fed Ex and internationally to Varian International Customer Support Managers for translation and notification of customers. The notice described the product, problem and actions to be taken. The customers were instructed to always acquire both the treatment port and the open port images BEFORE displaying the digital graticule, when using a standard double exposure template; and to
Quantity1486 devices
DistributionWorldwide distribution: USA (nationwide) and countries of: Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France,, Georgia, Germany,Greece, Hong Kong, Hungary
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K091209PORTALVISION ADVANCED IMAGING
Varian Medical Systems, Inc., Jun 23, 2009