Recall Z-1318-2012 ยท BSZ Gas-Machine, Anesthesia

Penlon, Ltd.

Anesthesia Machines (digital display assembly); A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation.

DateMar 26, 2012 (initiated Nov 29, 2011)
StatusTerminated
ReasonPenlon, Ltd has identified the potential for the premature failure of a component in some Electronic Flowmeter sensor display unit PCB assemblies.
Root causeComponent design/selection
ActionPenlon Limited sent a Technical Bulletin dated November 22, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. As a precaution a field upgrade part was released that required all existing PCB assemblies to be removed and the updated item installed. The old PCB must be returned to Penlon with the serial number of the affected machine. The required replacement parts were included with the technical bulletin. For further information customers should contact the Technical Support Team at techsupport@penlon.com.
Quantity160 units
DistributionWorldwide Distribution - USA (nationwide) and the countries of Peru, Uganda, Kenya, Malaysia, Poland, Saudi Arabia, United Arab Emirates, Philippines, Equador, and Thailand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061102PRIMA ANAESTHESIA MACHINE
Penlon , Ltd., Jul 28, 2006