ClearanceBrief
Recall Z-0656-2012 ยท BTO Tube, Tracheostomy (W/Wo Connector)

Smiths Medical ASD, Inc.

Bivona TTS Cuffed Neonatal Tracheostomy Tube Sizes 2.5 - 4.0 (With a Straight Neck Flange) Product Usage: This tube is intended to provide direct airway access for a tracheostomized patient for up to 29 days. It may be reprocessed up to 5 times for single patient reuse.

DateJan 19, 2012 (initiated Nov 28, 2011)
StatusTerminated
ReasonDifficulty disconnecting accessories from the Bivona Uncuffed Neonatal, Pediatric and Flextend Tracheostomy Tubes TR3 connector, which may result in a delay in initiating therapy or patient discomfort. If excessive force is used to detach the accessory, this may result in decannulation of the tube which could result in serious injury or death; especially if no replacement tracheostomy tube is imm
Root causeDevice Design
ActionSmith Medical sent an Urgent Field Safety Notice dated November 28, 2011, via U.S. Mail to all affected consignees. The notice identified the product, description of problem, and actions to be taken. This notice included a Customer Information Bulletin and Urgent Field Safety Notice Confirmation Form. Customers were instructed to 1) identify all affected unused product in inventory and segregate it to a quarantine location; 2) complete and return the attached Confirmation Form by fax to 219-989-7259 or email to bivona.tr3@smiths-medical.com; and 3) distributors were instructed to stop dist
QuantityUS distribution 153,033 devices ; Foreign 21,490 devices
DistributionWorldwide Distribution - USA Nationwide: AZ, CA, CO, CT,DE, FL, GA, HI, IL, IN, KS, KY, LA, SC, TX, OR, NJ, MO, UT, MI, NY, NM, MA, MN, MS, NC, ND, NE, NH, NV, OH, OK, OR, RI, SD, TN, VA, VT, WA, WI, WV including Washinton DC, Australia, Austria, Barbados, Belgium, Brazil, Canada, Chile, Cyprus
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K912469BIVONA WIRE REINFORCED PEDIATRIC TRACHEOSTOMY TUBE
Bivona Medical Technologies, Jun 19, 1991