Recall Z-0806-2012 ยท FEQ Pump, Air, Non-Manual, For Endoscope

Olympus America Inc.

Olympus OFP Flushing Pump irrigation tubing The Olympus OFP Flushing Pump is a peristaltic flushing pump intended for use with Olympus gastrointestinal videoscopes having an auxiliary water inlet as part of an Olympus endoscopic imaging system to facilitate washing of gastric and colonic mucosa to remove blood, feces and other organic matter, enabling improved visualization, diagnosis and therapy

DateJan 18, 2012 (initiated Dec 12, 2011)
StatusTerminated
ReasonWater may leak from the irrigation tubing used on the Olympus Flushing Pump
Root causeNonconforming Material/Component
ActionThe firm,Olympus, sent an "URGENT-DEVICE CORRECTION " letter dated December 12, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were informed that any current irrigation tubing in their inventory can be used with the OFP pump; however, they must inspect the tubing installed on the OFP pump for signs of water leakage; if leakage is found, discontinue use of the tubing and replace with a new irrigation tube. In addition, complete and return the enclosed questionnaire via fax to (484) 896-7128. If you do not wish to use your current inve
Quantity64,667 packages of 10 irrigation tubes
DistributionNationwide distribution: USA including states of: AL, AK, AZ, AR, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, AND WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K000948OLYMPUS ENDOSCOPIC FLUSHING PUMP MODEL OFP-1
Keymed (Medical & Industrial Equipment), Ltd., May 12, 2000