Recall Z-0680-2012 ยท LYZ Vinyl Patient Examination Glove

Cypress Medical Products LLC

Zee Clear Vinyl Medical Examination Gloves, Powder Free, Large, Non-Sterile; 100 gloves per box, 10 boxes per case; Distributed by Zee Medical, Inc., Irvine, CA 92606, Made in China; Zee Part No.: 3062 Single Use, medical exam glove worn on the hand to provide a barrier against potentially infectious materials and other contaminants

DateJan 11, 2012 (initiated Nov 15, 2011)
StatusTerminated
ReasonThe gloves fail to meet the performance requirements of 21 CFR 800.20 for leakage defects.
Root causeNonconforming Material/Component
ActionCypress Medical telephoned the accounts and then sent Urgent Voluntary Medical Device Recall letters dated November 14, 2011, to all affected customers. The letter informed them that the lot of gloves did not meet certain release specifications and that some of the gloves may not provide an effective barrier to infectious material. Customers were requested to cease distribution of the affected product, quarantine the gloves, and contact Cypress Medical's customer service department at 800-334-3646 for a returned goods authorization, return instruction and credit. The accounts were also requ
Quantity240 cases
DistributionNationwide Distribution including California, Indiana, New York and Texas
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093264VINYL EXAMINATION GLOVE, POWDER FREE, NON-COLORED
Qingdao Lang Hui Import & Export Co., Ltd., May 28, 2010