Recall Z-0594-2012 ยท IYE Accelerator, Linear, Medical

Philips Medical Systems (Cleveland), Inc.

Pinnacle3 Radiation Therapy Planning System, software version 9.0. Model # Description 453560446041 Pinnacle3 version 9.0 English 453560446051 Pinnacle3 version 9.0 Simplified Chinese 453560446061 Pinnacle3 version 9.0 Traditional Chinese 453560446091 Pinnacle3 version 9.0 Dutch 453560446101 Pinnacle3 version 9.0 French 453560446111 Pinnacle3 version 9.0 German 453560446141 Pinnacle3

DateJan 11, 2012 (initiated Nov 18, 2011)
StatusTerminated
ReasonPhillips Medical Systems have recently determined that a software (Pinnacle3 Software version 9.0) nonconformance can cause incorrect beam geometry.
Root causeSoftware design
ActionConsignees were sent on 11/18/11 a Philips "Urgent Medical Device Correction" letter dated 2011 Nov 16. The letter identified the affected product and described the Problem, Hazard Involved, How to Identify Affected Products, Action to be Taken by Customer/User, Actions Planned by Philips and Further information and Support. If customers observed the issue, they were to manually make the suggested change. In addition, customers were to place the letter in the Instructions for Use until otherwise notified. The firm will deliver 9.2 software release free of charge. Customers in North Americ
Quantity715 (463 US)
DistributionWorldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV, and countries of AUSTRALIA, AUSTRIA, BELGIUM,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041577PINNACLE3 RADIATION THERAPY PLANNING SYSTEM
Adac Laboratories, Jun 25, 2004