Recall Z-0744-2012 ยท FRN Pump, Infusion

Baxter Healthcare Corp.

Baxter Colleague CX Infusion Pump, Single Channel Volumetric Infusion Pumps. Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015. Product Code: 2M8161. Product Usage: Intended for use in a wide variety of patient environments for adult, pediatric and neonatal patients. Electronic infusion pumps indicated for continuous or intermittent delivery of solution through clini

DateJan 13, 2012 (initiated Jul 14, 2010)
StatusTerminated
ReasonBaxter's Service Center did not perform the ground impedance and leakage current test which verify if any power overflow occurs and that it is properly discharged through the device electrical system.
Root causeEmployee error
ActionBaxter sent service technicians to the consignees location on July 14, 2010 and October 6, 2010 and conducted the ground impedance and leakage current test on the affected pumps. The pumps were found with specifications.
Quantity2 units
DistributionNationwide distribution: Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063696COLLEAGUE VOLUMETRIC INFUSION PUMP MODEL# 2M8151,8153,8161,8163,9161,9163
Baxterhealthcare, Inc., Feb 27, 2007