Recall Z-0519-2012 ยท FQO Table, Operating-Room, Ac-Powered

Smith & Nephew, Inc. Endoscopy Division

Smith & Nephew, SPIDER2 Limb Positioner, including accessories Battery, foot pedal and charger Catalog Number: 72203299. Table, Operating-room, Pneumatic.

DateJan 11, 2012 (initiated Jul 26, 2011)
StatusTerminated
ReasonSPIDER2 may unlock (loss of traction) resulting in potential movement of the patient when the device is within close proximity to an RF source. Loss of traction can potentially result in the unintended repositioning of the extremity and potential for tissue or cartilage damage
Root causeDevice Design
ActionSmith & Nephew Advanced Surgical Devices Division sent an "URGENT- PRODUCT RECALL NOTIFICATION-URGENT" letter dated July 26, 2011 to all affected customers. The letter described the product, problem, and the actions to be taken by the customers. Customers were instructed to complete and return a form via fax at 405-917-8650. Contact the firm at 405-917-8556 for questions regarding this recall.
Quantity52 units
DistributionWorldwide Distribution-USA (nationwide) and the countries of Canada and France.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.