Recall Z-0596-2012 ยท NGT Saline, Vascular Access Flush

Hospira Inc.

0.9% Sodium Chloride Injection, USP, pre-filled flush solution, for IV Flush only, sterile fluid path; 10 mL Single-Use Syringe with Male Luer Lock, 100 syringes per pack, 4 packs per case; Made in Costa Rica, Hospira, Inc., Lake Forest, IL 60045 USA; list no. 1978-20 lock flush solution

DateJan 11, 2012 (initiated Nov 18, 2011)
StatusTerminated
ReasonThere is particulate matter in the fluid pathway, which has been identified as the same material as the rubber tip of the syringe plunger.
Root causeMaterial/Component Contamination
ActionHospira sent a "URGENT DEVICE RECALL" letter dated November 18, 2011, to all affected customers via UPS. The letter identified the product the problem and the action needed to be taken by the customer. The customers were instructed to check their inventory and immediately quarantine any affected product complete and return Reply Form via fax to 1-866-912-2512. Return affected product to Stericycle using the label provided with the letter and if they have distributed the product further, notify their accounts that received the product . Please contact Hospira Customer Care at 1-877-946-7747
Quantity5,817,600 syringes
DistributionNationwide distribution including states of: Alabama, Alaska, Arizona, California, Colorado, Connecticut, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Mississippi, Missouri, Nebraska, New Hampshire, New Jersey, New York, No
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090882FLUSH SYRINGE FILLED WITH 0.9% SODIUM CHLORIDE
Hospira, Inc., Apr 23, 2010