Recall Z-0342-2012 ยท IYE Accelerator, Linear, Medical

Elekta, Inc.

Elekta Synergy XVI X-ray Volume Imaging System To be used as part of radiation therapy treatment process.

DateDec 1, 2011 (initiated Oct 20, 2011)
StatusTerminated
ReasonA small number of sites have experienced situations with kV generator transformers overheating due to a diode failure on the interface board.
Root causeNonconforming Material/Component
ActionThe firm, Elekta, issued a "FIELD CHANGE ORDER MANDATORY ACTION" notice dated October 20, 2011 to its customers to add fuse protection for the kV transformers. The notice describes the product, problem, provides instructions and procedures and actions to be taken. The customers were instructed to follow any instructions or recommendations covered in the notice and to file the document in the Important Notice section of the appropriate User Manual. Elekta will contact the customer if the FCO is applicable to their kV generator. If you have any questions, call 770-300-9725.
Quantity123
DistributionNationwide distribution: USA including states of: AL, AK, AZ, CA, CT, FL, GA, IL, IA, KY, MD, MS, MI, MN, MS, MO, MT, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TX, VT, VA, WA, WV, WI, Puerto Rico, & DC.