ClearanceBrief
Recall Z-0414-2012 ยท GEX Powered Laser Surgical Instrument

Angiodynamics Worldwide Headquarters

VenaCure EVLT NeverTouch-FRS PROCEDURE KIT Featuring Gold-Tipped Fiber, 65 cm, VC NT FRS 65cm Clear Non-Echo Needle, Catalog No./REF 51403005, STERILE -- AngioDynamics, Inc. 603 Queensbury, NY USA 12804. This is a Special Order. Product Usage: The AngioDynamics, Inc. VenaCure Procedure Kit is indicated for endovascular coagulation of the Great Saphenous Vein (GSV) in patients with superficial

DateJan 11, 2012 (initiated Nov 10, 2011)
StatusTerminated
ReasonDuring an in-process inspection the hemostasis sidearm assemblies were identified to have the potential for flash to exist on the housing at the base of the threads. This flash has the potential to break off and relocate within the fluid path.
Root causeProcess control
ActionAngioDynamics sent an URGENT: Medical Device Recall letters and response form (dated November 10, 2011) to all affected customers. The letter identified the affected products, the problem, and the actions to be taken by customer/user. The customers were instructed to do the following: 1) Identify, segregate and return the recalled products to ANGIODYNAMICS and 2) Complete the enclosed Recall Reply Form even if they do not have any products remaining in their possession, noting zero quantity to be returned and fax to 518-798-1360. The letter states that a returned authorization number (RA#)
Quantity105 units
DistributionWorldwide Distribution - USA (nationwide) and the countries of: Canada, Great Britain, France, India, New Zealand, Netherlands, Saudi Arabia, Turkey, and Singapore
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102796ANGIODYNAMICS, INC. VENACURE PROCEDURE KIT
AngioDynamics, Inc., Dec 15, 2010