Recall Z-0816-2012 ยท FPO Stretcher, Wheeled

Sechrist Industries Inc

Sechrist Hyperbaric Chamber Gurney. The Sechrist Hyperbaric Chamber Gurney is an accessory device to the Sechrist Monoplace Hyperbaric Chambers. It is intended solely for the loading and unloading of patients into Sechrist Monoplace Hyperbaric Chambers. It is not intended for general hospital patient transport and is sold only to be used in conjunction with Sechrist Monoplace Hyperbaric Chambers.

DateJan 19, 2012 (initiated Sep 15, 2009)
StatusTerminated
ReasonStryker Medical has received complaints from customers that indicate a bushing in the braking system may fracture, which could result in the brake/steer pedal seizing. If the pedal seizes, the braking system may become inoperable or steering around corners may come more difficult. While no injuries have been associated with these reported events nor has there been reported failure events resulting
Root causeNonconforming Material/Component
ActionSechrist Industries sent a Urgent Medical Device Correction letter dated September 9, 2009, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1. Locate the gurneys with the serial numbers listed in this advisory. 2. If any of the affected gurneys cannot be located, determine if they have been sold, transferred, or donated to another facility and notify us regarding their new location. 3.Verify the braking system is working properly. If the brake system is inoperable, or if the brake and /or steer require a greater t
Quantity198
DistributionWorldwide Distribution -- USA (nationwide) and the countries of Peru, Venezuela, Brazil, Egypt, Mexico, Austrailia, Denmark and Taiwan