Recall Z-0839-2012 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Terumo Cardiovascular Systems Corporation
Roller Pump, 4 Inch Diameter The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
| Date | Jan 24, 2012 (initiated Jan 13, 2012) |
| Status | Terminated |
| Reason | Terumo Cardiovascular Systems (Terumo CVS) discovered and rejected a single digital encoder cable assembly during incoming inspection because it had a loose wire. The assembly is used in the 4" and the 6" roller pumps for the Terumo Advanced Perfusion System 1. |
| Root cause | Nonconforming Material/Component |
| Action | TERUMO sent an Urgent Medical Device Correction letter dated January 13, 2012, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to review the Medical Device Correction notice. Assure that all users are aware of the notice. Confirm receipt of the communication by faxing the attached Customer Response Form to 1-900-292-6551. Roller pumps with affected cables will have the cables assembly replaced. For questions regarding this recall call 1-800-521-2818. |
| Quantity | 44 |
| Distribution | Worldwide Distribution - USA including MD, WI, and IA and the countries of Canada, Mexico, and Vietnam |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K022947 | ADVANCED PERFUSION SYSTEM 1 (100/120 VOLT),ADVANCED PERFUSION SYSTEM 1 (220/240 VOLT, ROLLER PUMPS (6, 4 INCH Terumo Cardiovascular Systems Corp., Mar 20, 2003 |