ClearanceBrief
Recall Z-0951-2012 ยท HRX Arthroscope

Ascent Healthcare Solutions

***CAT 7205326***DYONICS***Arthroscopic Burr, Wide Mouth***4.0mm ACROMIONIZER, 13cm long***Fits Power/EP-1 Motor Drives***Mauve***SAMPLE LATEX-FREE***Made in the USA***Manufacturer: ASCENT HEALTHCARE SOLUTIONS***Lakeland, FL*** Lot number: 195769***Expiration Date: 11/2011. Reprocessed arthroscopic shavers are indicated for use in orthopedic surgical procedures of the joints, jaw or sinuses wher

DateFeb 3, 2012 (initiated Jan 26, 2009)
StatusTerminated
ReasonOn 01/26/2009, Ascent Healthcare Solutions, Lakeland, FL initiated a recall of various models of Dyonics, Stryker, and Linvatech Reprocessed Arthroscopic Shaver Blades. The sterility of the devices may be compromised due to a potential breach in the seal of the Tyvek cover to the plastic tray.
Root causePackaging process control
ActionAscent Healthcare Solutions sent an "URGENT MEDICAL DEVICE VOLUNTARY RECALL" letter dated January 26, 2009 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Customers were instructed to examine their inventory, remove and quarantine all affected devices. A Recall Effectiveness Check Form was attached for customers to complete and return via fax to 1-866-325-4760. Contact the firm at (888) 888-3433 ext. 5519 for questions regarding this recall.
Quantity36 devices
DistributionWorldwide Distribution-USA (nationwide) including the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, ND, NH, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, WA, WI, and WY, and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012611REPROCESSED DYONICS ARTHROSCOPIC SHAVERS
Alliance Medical, Inc., Nov 7, 2001
K012652REPROCESSED DYONICS ARTHROSCOPIC BURS
Alliance Medical Corp., Nov 8, 2001