Recall Z-0447-2012 ยท HBF Drills, Burrs, Trephines & Accessories (Compound, Powered)

Biomet Microfixation, Inc.

Box labeled in part, "REF: 72-1000***IQ SYSTEM IQ INTELLIGENT DRIVER 16.51CM (6.5")(L) RABALON, STAINLESS STEEL, GOLD PLATED***NON-STERILE PRODUCT*** MADE IN THE PHILIPPINES*** BIOMET MICROFIXATION 1520 TRADEPORT DRIVE JACKSONVILLE, FL 32218 USA www.biometmicrofixation.com***" The Biomet Microfixation IQ Intelligent Driver is a handheld cordless device designed for three functions: drill pilot ho

DateJan 11, 2012 (initiated Jul 8, 2011)
StatusTerminated
ReasonBiomet Microfixation, Inc., Jacksonville, FL, recalled their IQ Intelligent Device, 72-1000 due to failure of the front bit sensor to detect the correct bit that is installed. When a screw is inserted with the device in drill mode, it is possible for the screw or bone to strip.
Root causeComponent design/selection
ActionBiomet Microfixation called, emailed, or sent an Urgent Medical Device Recall letter dated July 8, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to remove the remaining device(s) at their facility and return immediately to PIVOT for the required software upgrade. Customers were also instructed to return the attached reconciliation form to: PIVOT International 10916 Strang Line Road Lenexa, KS 66215 Attn: Julie Lawrence For any questions regarding this recall call 1-800-874-7711 or 904
Quantity161 devices were distributed.
DistributionWorldwide Distribution - USA AZ, CA, CO, DC, FL, GA, HI, ID, IL, IN, KS, LA, NC, NH, NY, OK, OR, PA, TN, TX, and WI and the countries of Argentina, Brazil, France, Italy, Japan, South Africa, and Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.